ProductsYour current location:Home > Products > Record Monitor > 3 channel > CB-1303-C

CB-1303-C

CB series ambulatory ECG system is much more than a standard Holter analysis system. It combines advanced techniques to deliver unprecedented diagnostic capabilities.

Every detail from recorder design to system software includes techniques to deliver near perfect results. Leading technology, outstanding performance and functionality have made the CB series the most advanced user-friendly system available.

Traditional Holter analysis systems generally have two problems:

1. Data playback and analyzing speed is time consuming.

2. Manually modifying settings to try to achieve accuracy and repetitious editing can become tiresome, especially with cases with serious artifact. Accuracy in most cases is compromised.

The above problems directly effect efficiency and accuracy and produce low quality medical diagnostics. Often patients must return to have the test repeated causing serious inconvenience to your medical facility and the patient. With the rising costs of operation, today's successful health care providers must use cutting edge diagnostic equipment and services to remain profitable. The right technology and services determine the success of the health care provider and the overall experience and well being of the patient. Our products help enhance customer experiences and improve overall business performance and profitability. Our products help increase customer retention, reduce costs and develop new revenues.

The BIOX advanced edition Holter analysis system surpasses the competition in efficiency and accurate cardiac diagnostics. On average complete editing and report generations take under 10 minutes.

Advanced recorder

The "pager" sized durable mini-recorder can be put into your pocket. The compact design allows patients to go about their daily activities without being burdened by the utilization of the recorder.

Highly advanced integration design allows a compact conformation to improve reliability and is shock and vibration resistant.

Convenient single-button operation, accompanied by a three color LED/LCD for easy recognition of operation mode, extremely easy to operate

Advanced designed amplifier and A/D convertor circuits allows the signal fidelity index to exceed traditional standards. The sampling rate can be adjusted up to 10000Hz. The resolution can be as high as 1/4096, which not only allows a superior high quality ECG, but also reflects the changes in ST segment as well as SAECG.

Data storage utilizes advanced SD card technology; this technology allows for a more compact configuration for storage of all ECG data.

The designed structure ensures the SD card cannot be dislodged at any time. Unsuccessful recording is avoided.

International standard electrode placements (5 or 7 electrodes/3-channel, 10 electrodes/12-lead)

Independent pacemaker detection channel with highest sampling rate up to 10000Hz

Pre or Post record patient information into the system jointly with the patient diary, to provide auxiliary reference for the analysis of ECG

The wireless infrared interface/LCD provides real-time monitoring function. The quality of patient hook-ups can be previewed "live" to increase data quality and ensure valuable clean data, which effectively lowers artifact and reduces re-testing. In addition it can be utilized as a wireless ECG monitor.

Preset the patient's information into the recorder to avoid confusion, universal design suits facilities with multiple recorders.

The USB2.0 technology allows fast playback within 30 seconds, which greatly improves efficiency.

One AA or AAA size alkaline battery allows recording for 72 hours.

The BIOX software's structural technology allows seamless and endless potential to update future analysis. BIOX is continually updated and adding new analysis capabilities.

High-speed, accurate system software

Operator login system to prevent illegal entry, which ensures confidentiality of patient

Receive data in 30 seconds and can automatically analyze in 10 seconds.

QRS detection with four categories of wave-form recognition: ventricular, supra-ventricular, normal and artifact. Every QRS wave-form is categorized and marked to ensure the accuracy of the analyzed results.

Breakthrough technology applies a second order of analysis, this analyzing technique replaces heavy user interaction and complicated manual modification to reproduce accurate results. The automatic batch processing applies redefining classifications to every suspect wave-form. The revolutionary technique greatly improves the quality and accuracy.

Flexible and streamlined operation of editing QRS waves are viewed in singles, rows, lines or by complete category.

Determinant threshold values can be adjusted to meet the requirements of even the most complicated patient needs or clinical research.

The beat by beat editing tool applies redefinition, deleting and classification is quick and convenient.

The easy-to-operate measuring tools provide "gauge" functions.

Comprehensive ST segment report with each ST segment episode that includes deviation, slope, duration and heart rate.

Comprehensive analyzing functions. In addition to the advanced arrhythmia and ST segment analysis, the system includes heart rate variability (HRV) time domain, frequency domain and nonlinear analysis, pacemaker analysis, QTc analysis, atrial fib auxiliary analysis, sleeping respiratory analysis, SAECG time domain and frequency domain analysis, heart rate turbulence, T-wave alternate analysis* and much more.

The left button and right button of the mouse can be used jointly with the keyboard, making it convenient for experienced operators to work quickly and efficiently.

Browsing ECG data can be viewed in many modes, including standard ECG, ECG thumbnail and paging scanning. When browsed in ECG thumbnail and paging scanning, applied colors are used and can be redefined to denote normal, artifact, ventricular beat, supra-ventricular beat and pacing rhythm make it a breeze to interpret.

Abundant graphs, charts, statistical tables and category-specific histograms.

The diagnosis notes can be edited and printed on the summary page of the report.

The user has full control over the customization of the color, style, editing and printing of the report.

Print reports with your institution's logo and facility information.

Logical on-screen command buttons are easy to understand.

The multi-functional software analysis can process multiple tasks simultaneously. When you are editing or browsing, the systems software is calculating and adjusting all your changes and editing automatically.

The system can automatically generate various typical diagnostic strips with classification, time of occurrence, heart rate and beat annotations. You can conveniently choose to print any event or insert additional ECG-strip information.

Our comprehensive mouse-driven second wave template editing re-classifies all similar waveforms instantly

Sorting out QRS complexes is instant and categorized by color

QRS waves are color categorized, with a quick glance they can be verified instantly

3-channel ECG

Technical Specifications

Detecting channels:  

3 channels ECG

 Battery voltage recording channel

Independent pacemaker detection

 Independent SAECG detection(option)

Patient Input configuration:

5-lead (7-lead, option)

Input dynamic range:

±5 mV ECG, ±300 mV offset voltage

Input impedance:

> 10 MΩ

Common mode rejection:

≥ 80dB